Evaluating
the performance of digital mass-flow controllers
Mohamed
Saleem and Sowmya Krishnan, Ultra Clean Technology
With
the advent of 300-mm and single-wafer processing, the semiconductor
industry has placed increasing demands on precise process control and
repeatability. One area requiring stringent process control to meet
the demands of decreasing device geometries is the delivery of gases
to the process tools. Controlling gas delivery, in turn, requires accurate
flow control devices such as mass-flow controllers (MFCs).
Typically,
gas delivery is accomplished by means of an MFC located on a process
tool among other gas-delivery components. Most MFCs operate on the principles
of heat and mass transfer. Analog MFCs were previously the workhorse
of the semiconductor industry. However, despite their widespread use,
conventional analog MFCs have limitations in specific areas of gas-flow
range. Most analog MFCs become less accurate at flows below 10% of the
full-scale flow range. Their accuracy specification is normally a function
of full-scale flow, and flow errors are magnified when the MFC operates
at less than its specified full-scale flow. Most MFC manufacturers specify
accuracy in terms of ±1% of full scale for 2100% flow ranges.
For example, an analog MFC rated to perform at 1000 std cm3/min
has an error range of ±10 std cm3/min. If the unit operates
at a setpoint of 10% (100 std cm3/min), the flow rate is
100 ±10 std cm3/min; in other words, the flow rate has
a ±10% error range. The error is further magnified when the MFC
operates below a setpoint of 10%.
Thus,
while the flow accuracy of an analog MFC is about 1% near full scale,
it drops several percentage points at low flow rates (<25% of full-scale
flow). That loss of accuracy poses repeatability problems at low flow
rates, limiting the overall functionality of analog MFCs. Additionally,
analog MFCs cannot be reconfigured easily to accommodate changing flow
or gas conditions, as required by process engineers in the fab.
To
avoid the limitations of analog MFCs, digital MFCs are becoming increasingly
visible in the semiconductor industry. They enjoy several advantages
over analog MFCs, including improved accuracy over a wide range of setpoints
(generally ±1% accuracy over a setpoint range of 25100%)
and ease of calibration and configuration in multigas/multiflow applications
at end-user sites. By means of extensive modeling and gas characterization,
MFC manufacturers have been able to develop user-configurable units
that permit end-users to program them for virtually any gas or flow
rate within a specified boundary.1 That flexibility permits
manufacturers to reduce their inventories of MFCs, eliminates recalibration
costs arising from configuration changes, and minimizes or eliminates
cancellation costs associated with configuration changes.
Digital
MFCs have become increasingly accurate because they calibrate and store
data at multiple points throughout their operating range. Their accuracy
specification, based on percent of setpoint as opposed to percent of
full scale employed by analog MFCs, leads to improved linearity and
accuracy specifications.2
This
article discusses the results of an evaluation of several digital MFCs
from different manufacturers. Two digital MFCs (a 500std cm3/min
N2 unit and a 10std L/min N2
unit) from each manufacturer were tested for particles, moisture drydown
time, flow accuracy, and multiflow and multigas capability. While particle
performance and moisture drydown characteristics are not directly related
to the digital capabilities of MFCs, they are important for understanding
how far MFCs have advanced technologically.
Test
Protocol and Results
Particle
Characteristics. Each MFC was tested for particles using a Micro-LPC-HS
laser particle counter (LPC) from Particle Measuring Systems (Boulder,
CO). The range of particle sizes tested was 0.051.0 µm. The
LPC has a sampling rate of 2.8 std L/min and operates at atmospheric
pressure. A pressure diffuser was used to reduce the upstream pressure
and flow to the LPC. Particle tests were performed in both static and
dynamic modes. Static tests were performed for 30 minutes, during which
time the MFC remained fully open. Dynamic tests were performed for 10
minutes, during which time the MFC valve was cycled for 10 seconds every
minute. In other words, with the MFC at 100% setpoint, the unit's power
was turned off for 10 seconds and then turned back on again.
Particle
test data show that nearly all of the MFCs under investigation appeared
to be clean during 30 minutes of static testing in N2.
However, after undergoing 10 minutes of dynamic testing, a few MFCs
exhibited moderate particle shedding. Particle shedding was attributed
either to the internal design of the gas-wetted areas of the MFCs or
to contamination arising during assembly or packaging. In addition,
the MFCs used in this evaluation were test devices; they were not designed
to be used in volume production or to be installed in gas boxes.
Moisture
Drydown Characteristics. Moisture drydown tests were performed using
a UG-240A atmospheric pressure ionization mass spectrometer (APIMS)
from Hitachi Instruments (Tokyo). The tests were performed on 10-std
L/min MFCs only, since the APIMS requires 1.2 std L/min of flow for
testing. The MFCs were purged with nitrogen for 1248 hours until
their initial moisture level dropped below 1 ppb. To prevent compromising
their sensitive electronic circuitry, the MFCs were not baked to remove
hydrocarbon contaminants. The units were challenged with 200 ppb of
moisture and equilibrated for 20 minutes. Drydown was monitored by switching
to zero gas (200-ppt-purity nitrogen), which generated data indicating
moisture drydown performance after the device had been subjected to
a controlled moisture spike. The tests were performed according to SEMI
Standard F58-1000.
 |
|
Figure 1: Moisture drydown characteristics
for different MFCs.
|
The
MFCs' moisture drydown performance was evaluated using APIMS after a
200-ppb moisture challenge lasting 20 minutes. The results of that test,
shown in Figure 1, indicate that the different MFCs exhibited markedly
different moisture drydown capabilities. The time required for the devices
to reach a moisture level of 10 ppb ranged from 2 minutes in the best
case to 8 minutes in the worst. The time required for the MFCs to reach
the 1-ppb moisture level ranged from 62 minutes in the best case to
124 minutes in the worst. A few MFCs did not reach the 1-ppb level even
after 5 hours of testing. However, the SEMI drydown specification for
most gas panels is 20 ppb of moisture after 6 hours in dry nitrogen
at at flow rate of 1 std L/min. Therefore, it was expected that most
of the MFCs tested would meet current moisture drydown specifications.
Flow
Verification. All flow-verification tests were performed in nitrogen
using a Cal-Bench automated primary gas-flow calibration system from
Sierra Instruments (Monterey, CA). The system's flow range is 1 std
cm3/min to 50 std L/min and is based on primary measurements
of length and time as stipulated by the National Institute of Standards
and Technology. Tests for digital accuracy were conducted at 25, 50,
75, and 100% setpoints.
 |
|
Figure 2: Flow-accuracy plots
for 10-std L/min N2 MFCs tested in as-received digital
mode.
|
Of
the low-flow (500std cm3/min N2)
MFCs that were tested as received, nearly all met the accuracy specification
in the range of 25100%, although some of the MFCs were marginally
outside specification at the 25% setpoint.
Figure
2 shows the percentage error per setpoint for different high-flow (10-std
L/min N2) MFCs tested at 25, 50, 75, and 100% setpoints in
as-received condition in digital mode. The results for those MFCs were
much more scattered than the results for the low-flow devices. Less
than half of the 10-std L/min devices met accuracy specifications at
all setpoints. The remaining devices did not meet accuracy specifications
even at 50% setpoint. Those results clearly indicate a need for improvement,
especially at low setpoints.
MFC
Multiflow and Multigas Features
Unlike
analog devices, digital MFCs can be configured to accommodate multiple
flow rates (generally down to 3060% of the original full-scale
flow rate, depending on the type of MFC). The multigas and multiflow
capabilities of MFCs enable IC manufacturers to significantly reduce
inventories and to increase MFC mean time to repair or replace rates.
When
the gas-flow range of an analog MFC is changed, the device must be recalibrated.
Recalibration involves sending the device to the manufacturer for resizing
and testing. The resizing process results in increased turnaround times
and added production costs. In contrast, because digital MFCs have multiflow
capability, their flow range can be changed simply by changing their
flow limits, an operation that can be accomplished during system testing
or in the fab. In addition, fab personnel can use one MFC with a variety
of gases simply by reconfiguring the device. Most digital MFCs can be
programmed to accommodate a range of gases.
MFC
Multiflow Capabilities. To evaluate the MFCs' multiflow capabilities,
the investigators reprogrammed both low-flow (500std cm3/min
N2) and high-flow (10std L/min N2)
devices using the software provided by the MFC vendors to accommodate
the lowest flow rates. The reprogrammed flow rates of the low-flow MFCs
ranged from 280 to 325 std cm3/min. For example, one 500std
cm3/min MFC was reprogrammed to a new full-scale flow rate
of 280 std cm3/min N2.
 |
Figure 3: Flow-accuracy plots from
multiflow tests of low-flow MFCs in digital mode.
|
Figure
3 plots the percentage error versus setpoint for the low-flow MFCs.
The figure indicates that the percentage error was largest at the lower
setpoints for most MFCs and decreased to within acceptable limits at
the higher ones. Less than half of the MFCs tested met accuracy specifications
at all setpoints. A few MFCs were out of specification at 25 and 50%
setpoints. Most high-flow MFCs, on the other hand, did not meet specifications
at any setpoints, although some of them were only marginally off.
 |
Figure 4: Flow-accuracy plots from
multigas tests of low-flow MFCs in digital mode.
|
Multigas
Capabilities. MFCs with multigas capability can be configured for
any gas listed in SEMI's gas tables (SEMI Standard E52-0302). To test
that capability, the investigators reprogrammed both low- and high-flow
devices for argon gas. Figure 4 presents an error plot indicating that
all of the low-flow MFCs except one were within accuracy specifications
at all setpoints.
Figure
5 shows an error plot for high-flow MFCs that were tested after being
reprogrammed for argon gas. The results were consistent with the results
from the as-received and multiflow tests. Only one MFC met accuracy
specifications at all setpoints, while the others were outside specifications
at nearly all setpoints. That test demonstrated the performance limitations
of the MFCs under investigation.
 |
Figure 5: Flow-accuracy plots from
multigas tests of high-flow MFCs in digital mode.
|
Conclusion
Digital
MFCs have significant advantages over analog MFCs. They are more accurate
and easier to calibrate and reconfigure in multiflow/multigas applications
at end-user sites. They can also reduce fab inventories. These features
were demonstrably evident in the studies presented here. All of the
digital devices tested were more accurate than their analog counterparts.
However, many of the devices did not meet accuracy specifications at
setpoints ranging from 25 to 100% and in multiflow/multigas conditions.
Those limitations may be overcome by improving the modeling of gas characteristics
and control algorithms. While digital MFCs should be improved in some
areas, they are superior to analog MFCs. They can potentially reduce
costs and offer flexibility to meet the configuration requirements of
delivering gas to the process tool.
Acknowledgments
The
authors would like to thank Dave Celli and Chris Burkhart from Novellus
Systems for supporting the study described in this article. They would
also like to acknowledge the MFC suppliers Advanced Energy Industries,
Brooks Instrument, Mykrolis, MKS Instruments, Unit (now Celerity), and
Horiba STEC for their participation in this program.
References
1. CF
Drexel, "Digital Mass Flow Controllers Come of Age," Solid State
Technology 39, no. 11 (1996): 99106.
2. P
Rudent, "Calibration Method and New Developments for High-Performance
Mass Flow Controllers," Semiconductor Fabtech (8th ed.): 167170.
Mohamed
Saleem, PhD, is a senior technology development engineer at Ultra
Clean Technology (Menlo Park, CA). An active participant in the
standards activities of gas-panel, corrosion, surface-analysis, and
stainless-steel task forces of SEMI, Saleem has published several papers
in the field of materials science and has presented many papers at major
conferences. He received an MS in chemical engineering from Tufts University
in Medford, MA, and a PhD in materials science and engineering from
the University of Florida in Gainesville. (Saleem can be reached at
650/323-4100 or msaleem@uct.com.)
Sowmya
Krishnan, PhD, is CTO of Ultra Clean Technology, where she oversees
the technology and product development areas for semiconductor gas-
and liquid-delivery systems. She has chaired sessions on contamination-free
manufacturing and gas delivery at Semicon West, Semicon Southwest, and
CleanRooms West. She has also chaired the standards task force at SEMI.
She has authored several technical papers and editorials in the fields
of microcontamination and semiconductor processing. She received MS
and PhD degrees in chemical engineering from Clarkson University in
Potsdam, NY. (Krishnan can be reached at 650/323-4100 or skrishnan@uct.com.)